PRODUCT DETAILS
PPRODUCT DETAILS
THR-150YDC

THR-150YDC

0.00
0.00
  
1.The steam can't go through to chamber when the door haven't close,and the pilot off inform.You can processing dry operate.
2.When the pressure inner over 0.027MPa,the door should be locked and can't be open.
3.If you not choose the sterilzing time and star the machine,buzzer sound.
4.If you mistake open the door in the course of start,the jacker value suto off and open the chamber value.
5.Lack water the poilt on,heating and sterilize poilt off,the buzzer sound and release steam.
6.Overload and leakage protection,when leakage or over current occur,it auto off the power.
7.Temperature controlled not work and the inner pressure over 0.165MPa,the safety value auto start open.

Specifications:
.
Model:THR-YDC
THR-150YDC
THR-200YDC
THR-280YDC
Technical Data:



Sterilizing Volume:
150L(dia480*850mm)
200L(dia515*1000mm)
280L(dia600*1000mm)
Working Pressure:
0.22MPa
0.22MPa
0.22MPa
Working Temperature:
134°C
134°C
134°C
Adjustable Temperature:
40°C-134°C
40°C-134°C
40°C-134°C
Heat Average:
≤±2°C
≤±2°C
≤±2°C
Time of drying:
0-60mins
0-60mins
0-60mins
Time of sterilization:
0-60mins
0-60mins
0-60mins
Power:
6KW/380V 50Hz
9KW/380V 50Hz
12KW/380V 50Hz
Overall Size:
600*1400*1300mm
670*1400*1650mm
770*1400*1780mm
Packing Size:
800*1650*1600mm
860*1650*1900mm
960*1650*2000mm
G/N.W
270/360KG
290/380KG
340/440KG