Syringe Pump |
Parameter |
Syringe Size | 5ml,10ml, 20ml, 30ml, 50ml ( optional: 1/2/3mL) |
Flow Rate | 0.0-150ml/h (0.1ml/h increment) 0.0-300ml/h (0.1ml/h increment)
0.0-600ml/h(o.1ml/h increment)
0.0-900ml/h (o.1ml/h increment)
0.0-1200ml/h (0.1mI/h increment, 1ml/h increment if over 1000ml/h) 50ml syringe
|
Precision | ≤±2%(mechanism≤±1%) |
Inquiry of total volume | 0.0ml-9999ml(0.1ml increment, 1ml increment if over 1000ml) |
vTBI | 0.0ml-9999ml(0.1ml increment, 1ml increment if over 1000ml) |
occlusion value | High(H) :800±200mmHg ( 106.7±
26.7kPa) Centre ( c) : 500±100mmHg ( 66.7± 13.3kPa) Low ( L) : 300±100mmHg ( 40.7±13.3kPa)
|
Alarm | Remind alarm,Occlusion alarm ,Nearempty alarm,Empty alarm,syringe barrel unlock alarm, syringe plunger disengage alarm,syringe flange disengage alarm, AC power failure,
Low battery alarm, Battery empty alarm,VTBtfinish reminder, infusion tube disconnect(option)
|
Power | AC :100v-240. 5OHz-60Hz Built-in Battery : DC15V Ni-Mh battery.Built-in battery can run for 10 hours at 5mI/h.
( It can be replaced by lithium ion battery pack if necessary.)
DC : DC15v. vehicle power
|
Running Environment | Temperature : +5C-+40-c Relative Humidity : 20%-90%
Atmosphere Pressure : 860hPa-
|
Transport And Storage Environment | Temperature : -20C-+55C Relative Humidity : 10%-93%
Atmosphere Pressure : 500hPa-1060hPa
|
Electric Safety Classification | Class 1, CF type ,Built-in battery , Continuous operation equipment |
water Proof Level | IPX4 |
Net Weight | 2.2kg(include fastener) |
Dimension | 320mm (L)×132mm ow)× 138mm(H) |
Function |
Kvo rate | Standby function |
Safety lock for PURGE/BOLus | Safety lock for flow rate adjust |
All brands of syringes are available | Automatic syringe size detection |
Preset occlusion detection pressure level | Bolus release |
Preset Volume To Be Infused | Inquiry of total infused volume |
Varia video and audio alarm | Bolus release |
RS232 interface | vehicle Power |
4500 history log | Double CPU |