Parameter |
Syringe Size | 5ml,10ml, 20ml, 30ml, 50ml ( optional: 1/2/3mL) |
Flow Rate | 0.0-150ml/h(0.1ml/h increment) 5ml syringe
0.0-300ml/h (o.1ml/h increment) 10ml syringe
0.0-600ml/h (o.1ml/h increment) 20ml syringe
0.0-900ml/h (o.1ml/h increment) 30ml syringe
0.0-1200ml/h (0.1ml/h increment, 1ml/h increment if over 1000ml/h) 50ml syringe
|
Precision | ≤±2%(mechanism≤±1%) |
Inquiry of total volume | 0.0ml-9999ml(0.1ml increment, 1ml increment if over 1000ml) |
VTBI | 0.0ml-9999ml(0.1ml increment, 1ml increment if over 1000ml) |
Occlusion Value | ±200mmHg ( 106.7±26.7kPa) Centre ( C) : 500±100mmHg ( 66.7± 13.3kPa) Low( L) : 300±100mmHg ( 40.7± 13.3kPa)
|
Alarm | Remind alarm,Occlusion alarm ,Near empty alarm, Empty alarm, syringe barrel unlock alarm, syringe plunger disengage alarm,syringe flange disengage alarm, AC powerfailure,
Low batteryalarm, Batteryempty alarm,VTBt finish reminder, infusion tube disconnect(option)
|
Power | AC : 100V-240V±10%. 50Hz-60Hz± 1Hz Built-in Battery : DC15V Ni-Mh battery.Built-in battery can run for 10 hours at 5ml/h.(It can be replaced by lithium ion battery pack if necessary. ) DC : DC15v. vehicle power
|
Running Environment | Temperature : +5C-+40°℃ Relative Humidity : 20%-90%
Atmosphere Pressure :860hPa-1060hPa
|
Transport And Storage Environment | Temperature : -20°C-+55C Relative Humidity : 10%-93%
Atmosphere Pressure : 500hPa-1060hPa
|
Electric Safety Classification | Class I, CF type , Built-in battery , Continuous operation equipment |
water Proof Level | IPX4 |
Net Weight | 3.5kg(include fastener)v |
Dimension | 320mm (L) × 193mm (w) × 138mm (H) |
Function |
kvo rate | Standby function |
Safety lock for PURGE/BOLuS | Safety lock for flow rate adjust |
All brands of syringes are available | Automatic syringe size detection |
Preset occlusion detection pressure level | Bolus release |
Preset Volume To Be Infused | Inquiry of total infused volume |
Dual channel | Power indicator |
RS232 interface | Vehicle Power |
4500 history log | Double CPU |