PRODUCT DETAILS
PPRODUCT DETAILS
THR-850

THR-850

0.00
0.00
  


The anesthesia machine sends anesthesia into the patient's alveoli through a mechanical circuit to form a partial pressure of the anesthetic gas. After dispersing into the blood, it directly inhibits the central nervous system, thereby producing the effect of general anesthesia.

Features:

  • Standard integrated metallic modular respiratory circuit; completely solves the pro6blem of gas sealing under high temperature disinfection, easy to clean and disinfect.

  • High precision five tube flowmeter, fitted with interlocking oxygen-nitrous oxide supply, guarantees oxygen output is never lower than 21%.

  • Nonhazardous environmentally friendly frame, precise construction, elegant design, equipped with automated self-checking function during startup and an automatic self-calibrating sensor.

  • Advanced and stable Adjustable Pressure Limiting (APL) valve, suitable for low and ultra low flow anesthesia.

  • Equipped with advanced international rotating Na-lime tank, easy-operate, requires only one hand, and uses imported PPSU material and temperature resistant to 134°C.

  • Multi-parameter Patient Monitor: Optional.


Specification:

Physical specifications 
Screen: 10.4 ''  LCD  Display Screen
SuitableAdult & Child
 Mode:pneumatically powered and electrically controlled system
Working Mode:Closed; Semi-Closed; Semi-Open
CircuitBreathing circuit integrated standards
Flowmeters:5 Tubes Flowmeters:  O2:0.1~10L/Min,  N2O:0.1~10L/Min; Air: 0.1~10L / Min
Trolley: Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions
Drawer unitTWO drawers  totally extractable
Technical Specifications
Gas requirement: Medical oxygen and nitrous oxide with a pressure ranging from O2: 0.32~0.6MPa ;   NO2: 0.32 MPa to 0.6 MPa.
safety valve <12.5  kPa
Respiratory Rate1~99bpm
oxygen concentration in the mixed gas N2O/O> 21%
Oxygen Flush: 25~75 L/min
modes of ventilationA/C, IPPV, SIPPV, IMV, SIMV, PCV, VCV, PEEP, MANUAL, SIGH
PEEP:0 ~ 2.0 kPa
Inspiratory trigger pressure-1.0kPa ~ 2.0 kPa
I/E Ratio:8:1 ~ 1:10, Has the inverse ratio ventilation
Tidal Volume0 ~ 1500 ML
Inspiratory Plateau:0~1s
O2 Concentration:21%~100%
Sigh: one deep breath per 70 ~ 120 controlled respirations, the inspiration time is 1.5 times of the setting point
Maximum safety pressure:≤ 12.5 kPa
Pressure limit range:0 ~ 6.0  kPa
Airway pressure alarm: Audible and visual and with yellow and red color indicatingLower: 0.2kPa~5.0kPa;   Upper: 0.3~6.0  kPa
±0.2 kPa
Tidal volume alarm: Audible and visual and with yellow and red color indicatingupper alarm: 50 to 2000ml, lower alarm: 0~1800ml
Oxygen concentration alarm: Audible and visual and with yellow and red color indicatingupper alarm: 21% ~100%;  lower alarm:10%~80%
Power Supply AlarmAc/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s

The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.
 
Operating conditions
Ambient temperature: 10 ~ 40℃
Relative humidity:no higher than 80 %
Atmospheric pressure:860 hPa ~ 1060 hPa
Power requirement: 100-120 Vac, 50/60 Hz;  
Attention: the AC power supply used for the anesthesia machine must be well   grounded.
Storage
Ambient temperature: -15℃ ~ +50℃
Relative humidity: no higher than 95 %
Atmospheric pressure: 86 kPa ~ 106 kPa.

It should be stored in a room without corrosive gas and well ventilated
Package
packaging boxcomply with the requirement of GB/T 15464

Between the packaging box and product, soft material with appropriate thickness provided to prevent loosening and mutual friction during transportation

Moisture protection and rain protection to ensure that product is protected from natural damage. 
Safty & Alarm
Oxygen AlarmIt alarms when the oxygen  supply from pipe or cylinders lower than 0.2MPa
Ventilation Volume AlarmLower: 0~12L/Min;   Up: 18L/ Min
Power AlarmIt Alrmas while AC and DC supply failure ; Keep alraming  time: >120s
Air Tract Pressure AlarmLower: 0.2kPa ~5.0 kPa;  Up: 0.3kPa ~ 6.0kPa