PRODUCT DETAILS
PPRODUCT DETAILS
THR-MJ-560B1

THR-MJ-560B1

0.00
0.00
  

Standard Configuration:

one High-precision evaporator, either Halothane, Enflurane,Isoflurane or Sevoflurane configuration.

Double-tubes O2 flowmeter, standard loop system, ABS Engineered plastic frame, children or adult bellows, screen display ventilator (IPPV, MANUAL, ventilation volume, tidal volume,respiratory rate) , real-time display of respiratory duct pressure, clear at one glance)

Optional: CO2 concentration monitoring

THR-MJ-560B1is an economic type of  anesthesia cart .

1) The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose one kind of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.

2) Reliable pneumatic transmission system.

3) Economic and proper configuration of technologies, optional as required by the Customer.

4) Modularized design and upgrading. Upgrading is available as required by the Customer.

Part of Anesthesia

Main body                      High-strength engineering plastic rack, light,

beautiful and corrosion resistant

Scope of application adult, child
Gas sourceO2, N2O: 0.27~0.55MPa
Flow meterO2: 0.1~1.0L/min   1.1~10L/min
Flow rate of rapid oxygen supply25~75L/min
Low oxygen pressure alarmThere will be sound alarm when the oxygen

pressure < 0.2MPa

Vaporizer it has the function of automatic compensation based on pressure,temperature,and flow rate. The regulation range of evaporator concentration is 0~5 vol%.

Among Halothane,Enflurane,Isoflurane and Sevoflurane,one can be chosen for application as required by the customer.

Respiratory circuitworking mode: all-close, semi-close, semi-open

releasingpressure: 0.5~7 kPa

Respiratory bellows   bellows for adults, bellows for children

tidal volume range:0~1500 ml

     

Part of Ventilator

Display modeHigh-definition 4.8" LCD screen display
Ventilation modeIPPV, MANUAL (Manual monitoring of tidal volume, ventilation volume, respiratory rate)
Ventilation parametersTidal volume         50~1500ml
Rate               4~80 bpm
I:E              2:1~1:4
Pressure range      1.0~6.0 kPa
Parameters for ventilation monitoringtidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway,inspiratory trigger pressure, pressure - time waveform.

Security Alarm System

Airway pressure alarmupper limit setting range 1.0~6.0 kPa

low limit setting range   0.4~2.0 kPa

TV alarmupper limit setting range 0.5~2.0L

low limit setting range  0.02~1.0L

Sustained high-pressure alarmit will give alarm when stress have consistently

been higher than 2.5 kPa

Suffocation alarm it will give sound and light alarm if there is

no tidal volume input in 15s

Power alarm
Power AC 110-220V±10%   50-60Hz±1Hz