PRODUCT DETAILS
PPRODUCT DETAILS
THR-MJ-S6100D

THR-MJ-S6100D

0.00
0.00
  

Specifications

  1. Adopts the electric control gas driven mode.

  2. With a variety of alarm, presentation and protection function

  3. Use 7 "color LCD display shows ventilation parameters, alarm information, and the graphical display.

  4. For adults and children

  5. O2, N2O flowmeter has the linkage,ensure O2 concentration≥25%;

  6. With O2 drive insufficient pressure alarm and N2O truncation protection

 

Working environment

  1. Air source     O2,N2O

  2. Air pressure                  280 kPa 600 kPa.

  3. Power supply 100-240V~

  4. Frequency 50 /60HZ

  5. Input power 40 VA

 
Ventilation mode

  1. Intermittent positive-pressure ventilation (IPPV)

  2. Assist / Control (A/C)

  3. Synchronized Intermittent Mandatory Ventilation (SIMV)

 

Main technical parameters

Adjustable parameters

  1. Flow meter

O2            0.1~10 L/min
N2O         0.1~10 L/min
O2 and N2O have internal linkage devices, ensure O2 concentration 25% 

  1. Oxygen flush                   35 L/min~ 75 L/min

  2. Tidal volume(Vt)                0 mL ~ 1500 mL

  3. MV                   For adult≥18 L/min                                                                            

  4. Frequence(Freq)          4 bpm 40 bpm

                                 (SIMV mode:1 bpm20 bpm)

  1. I/E                            2:1 ~ 1:6

  2. Press triggering sensitivity(PTr)-20 cmH2O ~ 0 cmH2O(Based on PEEP)

  3. Sigh Ventilation (Sign)            0(off), 1 ~ 5 (/100)

  4. Pressure Limit   20 cmH2O ~ 100 cmH2O

 
 
Monitored Parameters

  1. Frequence(Freq)          0 /min ~ 100 /min

  2. Tidal volume(Vt)                   0 mL ~ 2000 mL

  3. MV                     0 L/min ~ 100 L/min                                                                        

  4. airway pressure                     0 cmH2O ~ 100 cmH2O

 
Graphical display:

  1. P-T (Airway Pressure -- Time Waveform)

 
Alarm and protection 

  1. The AC power outage alarm

  2. No tidal volume

  3. High (low) ventilation alarm

  4. High (low)Airway pressure alarm

  5. Pressure limit

  6. Continuous pressure alarm  

  7. The maximum limited pressure  <12.5 kPa

  8. With O2 drive insufficient pressure alarm and N2O truncation protection

  9. O2, N2O flowmeter has the linkage to ensure O2 concentration≥25%.When O2 closed ,N2O closes at the same time.

 
Vaporizer 
 

Anesthesia gasAdjustable scope % (volume percentage)
Halothane0 ~ 5
Enflurane0 ~ 5
Isoflurane0 ~ 5
Sevoflurane0 ~ 8