Specifications
Adopts the electric control gas driven mode.
With a variety of alarm, presentation and protection function
Use 7 "color LCD display shows ventilation parameters, alarm information, and the graphical display.
For adults and children
O2, N2O flowmeter has the linkage,ensure O2 concentration≥25%;
With O2 drive insufficient pressure alarm and N2O truncation protection
Air source O2,N2O
Air pressure 280 kPa ~ 600 kPa.
Power supply 100-240V~
Frequency 50 /60HZ
Input power 40 VA
Ventilation mode
Intermittent positive-pressure ventilation (IPPV)
Assist / Control (A/C)
Synchronized Intermittent Mandatory Ventilation (SIMV)
Flow meter
O2 0.1~10 L/min
N2O 0.1~10 L/min
O2 and N2O have internal linkage devices, ensure O2 concentration 25%
Oxygen flush 35 L/min~ 75 L/min
Tidal volume(Vt) 0 mL ~ 1500 mL
MV For adult≥18 L/min
Frequence(Freq) 4 bpm ~ 40 bpm
(SIMV mode:1 bpm~ 20 bpm)
I/E 2:1 ~ 1:6
Press triggering sensitivity(PTr)-20 cmH2O ~ 0 cmH2O(Based on PEEP)
Sigh Ventilation (Sign) 0(off), 1 ~ 5 (/100)
Pressure Limit 20 cmH2O ~ 100 cmH2O
Monitored Parameters
Frequence(Freq) 0 /min ~ 100 /min
Tidal volume(Vt) 0 mL ~ 2000 mL
MV 0 L/min ~ 100 L/min
airway pressure 0 cmH2O ~ 100 cmH2O
Graphical display:
P-T (Airway Pressure -- Time Waveform)
Alarm and protection
The AC power outage alarm
No tidal volume
High (low) ventilation alarm
High (low)Airway pressure alarm
Pressure limit
Continuous pressure alarm
The maximum limited pressure <12.5 kPa
With O2 drive insufficient pressure alarm and N2O truncation protection
O2, N2O flowmeter has the linkage to ensure O2 concentration≥25%.When O2 closed ,N2O closes at the same time.
Vaporizer
Anesthesia gas | Adjustable scope % (volume percentage) |
Halothane | 0 ~ 5 |
Enflurane | 0 ~ 5 |
Isoflurane | 0 ~ 5 |
Sevoflurane | 0 ~ 8 |