Model | THR-LF06 | ||
External Size(L*W*H)mm | 2100*710*1900 | ||
Reagent Preparation Area | Operating Area Size(L*W*H)mm | 640*570*510 | |
Pressure | Positive | ||
Airflow Velocity(m/s) | 0.45 | ||
Airflow Volume(m3/h) | 420 | ||
Exhaust Type | - | ||
Specimen Preparation Area | Operating Area Size(L*W*H)mm | 660*570*510 | |
Pressure | Negative | ||
Airflow Velocity(m/s) | Downflow: 0.33±0.025 Inflow: 0.53±0.025 | ||
Airflow Volume(m3/h) | 250 | ||
Exhaust Type | TypeA2, 30% exhaust, 70% recirculation | ||
Amplification Analysis Area | Operating Area Size(L*W*H)mm | 640*570*510 | |
Pressure | Negative | ||
Airflow Velocity(m/s) | 0.4 | ||
Airflow Volume(m3/h) | 180 | ||
Exhaust Type | 100% exhaust | ||
HEPA Filter | Four, 99.999% efficiency at 0.3μm. Filter life indicator. | ||
Clean Level | Class 100(ISO Class 5) | ||
Front Window | Manual, Toughened glass, Anti UV | ||
Max Opening | 380mm | ||
Noise | ≤67 dB(A) | ||
Vibration | ≤5μm | ||
UV Lamp | 15W*3, UV timer, UV life indicator, emission of 253.7 nanometers for most efficient decontamination | ||
LED Lamp | 8W*3 | ||
Illumination(lx) | ≥650( Each area) | ||
Sockets | The load of two sockets (Each area): 500W, total 6 sockets | ||
Display | LCD | ||
Control System | Microprocessor | ||
Visual and Audio Alarm | Filter replacement, window over height, abnormal air flow velocity | ||
Work Surface Height | 750 mm | ||
Material | Work Zone: 304 stainless steel Main Body: Cold-rolled steel with anti-bacteria powder coating | ||
Power Supply | AC220V±10%, 50/60Hz; 110V±10%, 60Hz | ||
Consumption | 500W | ||
Standard Accessory | LED lamp*3, UV lamp*4, Windproof plate*2, Waterproof sockets*6 | ||
Package Size (W*D*H) | 2240*970*1450mm | ||
Gross Weight | 450kg |
Area | Name | Quantity |
Reagent preparation area | Pipette | 1 |
Pipette | 1 | |
Pipette | 1 | |
Centrifuge | 1 | |
Mixer | 1 | |
Specimen preparation area | Pipette | 1 |
Centrifuge | 1 | |
Mixer | 1 | |
Amplification analysis area | QPCR | 1 |