PRODUCT DETAILS
PPRODUCT DETAILS
THR-150YDC

THR-150YDC

0.00
0.00
  

Specifications:

-Sterilizing Volume: 280L(dia600*1000)mm

-Working Pressure 0.22MPa
-Working Temperature 134°C
-Adjustable Temperature 40 °C-134 °C

-Heat Average ≤±2 °C
-Time of drying 0-60mins
-Time of sterilization 0-60mins
-Power 12KW/380V,50Hz
-Overall Size 770*1400*1780mm
-Packing Size 960*1650*2000mm
-G/N.W:340/440KG

 
Description:
1.The steam can't go through to chamber when the door haven't close,and the pilot off inform.You can processing dry operate.
2.When the pressure inner over 0.027MPa,the door should be locked and can't be open.
3.If you not choose the sterilzing time and star the machine,buzzer sound.
4.If you mistake open the door in the course of start,the jacker value suto off and open the chamber value.
5.Lack water the poilt on,heating and sterilize poilt off,the buzzer sound and release steam.
6.Overload and leakage protection,when leakage or over current occur,it auto off the power.

7.Temperature controlled not work and the inner pressure over 0.165MPa,the safety value auto start open.

Technical data

 Models

THR-150YDC

THR-200YDC

THR-280YDC

THR-400YDC

Sterilizing chamber 

Volume

150L

φ480×850mm

200L

φ515×1000mm

280L

φ600×1000mm

400L

φ700×1100mm

working pressure

0.22MPa

working  temperature

134°C

Adjustment of temperature

40°C- 134°C

Time of sterilization

0-99 min

Time of drying

0-99 min

Heat average

≤±2°C

Power

9KW/380V 50Hz

9KW/380V 50Hz

12KW/380V 50Hz

18KW/380V 50Hz

Dimension

1400×600×1300

1400×670×1650

1400×770×1780

1430×880×1830