Conventional model | 15 d/ml infusion set 0.0ml/h~1200ml/h 20 d/ml infusion set 0.0ml/h~1200ml/h 60 d/ml infusion set 0.0ml/h~300ml/h Note:minimum resolution of setting flow rate is 0.1mL/h when rate is above 1000mL/h, the Resolution is 1mL/h. | ||
Drip Mode | 15 d/ml infusion set 0 d/min~300 d/min 20 d/ml infusion set 0 d/min~400 d/min 60 d/ml infusion set 0 d/min~300 d/min Note:The minimum resolution dripping speed Settings 1d/min. | ||
Accuracy | ≤±2%(Note:Under the condition of the benchmark test, using high elastic silicone iv path accuracy after calibration and commissioning.) | ||
Bubble Detector | Ultrasonic detection method:Bubble above 25ul is detected | ||
Total Infusion Volume | 0.0ml-9999ml(0.1ml/h increment,1ml/h increment if over 1000ml/h) | ||
Volume Limit | 0.0ml-9999ml(0.1ml/h increment,1ml/h increment if over 1000ml/h) | ||
Occlusion | (H):800±200mmHg (106.7±26.7kPa) (C):500±100mmHg (66.7±13.3kPa) (L):300±100mmHg (40.7±13.3kPa) | ||
History Record | These serial products can maximally store 4500 pieces of history records. The record includes the following information: Date, flow rate, alarm data, total volume delivered, occlusion pressure and syringe brand and serial number. | ||
Alarm | Near empty alarm,Empty alarm,Door open alarm, Bubble alarm,limited value reached alarm,AC power failure,Low battery alarm,Battery empty alarm,Remind alarm, Flow regulator or inlet valve closed alarm (optional) | ||
Power | Supply Voltage:AC100V~240V(±10%),Power Frequency: 50Hz-60Hz Built-in Battery:DC15V Ni-MH battery(It can be replaced by lithium ion battery pack necessary.) Working time of battery:In case the battery is fully charged, the continuous working time can be up to 6 hours at the speed of 25ml/h. Maximum power:30VA;Note:Before the first use of this product, the battery must be charged while being shut down for more than 12 hours. | ||
Working Environment | Temperature:+5℃-+40℃ Relative humidity:20%-90% Environmental pressure:860hPa-1060hPa | ||
Transport And Storage Environment | Temperature:-20℃~+55℃; Atmosphere Pressure:500hPa~1060hPa. products should be stored in the relative humidity in the range of 10% ~ 93%, no corrosion gas indoor and ventilated. | ||
Electric Safety Classification | Class I, CF type,Built-in battery,Continuous operation equipment | ||
Water Proof Level | IPX4 | ||
Net Weight | 2.2kg(include fastener) | ||
Dimension | 320mm(W)×132mm(H)×138mm(D) |