-32000 Hz Sampling Rate and 24 Sampling Data of each channel
-1 Second Baseline Baseline Stability Time
-Broad Frequency Response 0.01~250Hz
-Can Easily Capture Faint Dual-chamber Pacing Signal
-Utilize the University of Glasgow ECG analysis algorithm, which can automaticly give
-ECG interpretation based on age, gender, race, medication, and class.
-ST segment elevation myocardial infarction diagnosis ability is in the industry leading position.
-Support advanced analysis function like VCG, Time VCG etc
Auto analysis validation results | ||||||||
P-wave time limit (ms) | PR Interval (ms) | QRS Complex (ms) | QT Interval (ms) | |||||
Average Deviation | Standard Deviation | Average Deviation | Standard Deviation | Average Deviation | Standard Deviation | Average Deviation | Standard Deviation | |
ECG1200 results | 0.5 | 9.3 | -1.9 | 7.0 | 0.1 | 4.9 | 4.5 | 12.2 |
IEC60601-2-51 latest requirement | ±10 | 15 | ±10 | 10 | ±10 | 10 | ±25 | 30 |
Comparison of 601 male myocardial infarction patients diagnosis | ||
Age Group | ECG1200 automatic diagnostic effective rate | Cardiologists diagnostic effective rate |
18-39 | 100% | 87% |
40-49 | 96% | 87% |
50-59 | 93% | 90% |
60-69 | 93% | 82% |
70-79 | 92% | 91% |
≥ 80 | 91% | 93% |